Series 110 - Self-Cleaning Sampling Probe

Description

The sampling probe is designed to extract representative gas sample for stationary continuous emission monitoring system (CEMS) and process gas monitoring. It consists of a probe, a heated filter and an electrical module. The equipment removes particulate and evaporates any liquid mist before forwarding the gas sample to the sampling line. Depending on the application, automatic blow-back can be used for probe cleaning. This design ensures sample integrity and protects the monitoring system against contamination and corrosion.

Applications

The sampling probe meets all compliance monitoring requirements prescribed by US EPA Code of Federal Regulations for CEMS (40CFR60 & 40CFR75) as well as other relevant codes and standards. It can be used for sampling process gases and air emission sources, such as : power boilers, recovery boilers, incinerators, fluidized bed combustors, cement kilns, refineries, petrochemical processing facilities, drying or roasting kilns, scrubber exhaust, and others.

Self-Cleaning Function

An automatic blow-back facility can be utilized for probe cleaning. At a selected time interval, instrument air (steam optional) is discharged to blast the particulate and condensate back into the stack. Blow-back duration and frequency are user adjustable. The blowback operation can be activated by the built-in controller, or triggered by external control (such as AEM Model 120 Automatic Air Monitoring Controller, PLC, etc.). This feature ensures reliable extended continuous operation and minimizes maintenance.

Probe

Temperature and chemical resistant material are used to ensure that there is no sample loss. Depending on the application, the probe can be heated or cooled as required (optional).

Rodding Port

During normal operation, the probe can be clean using self-cleaning blow-back. To remove any build-up after extended operation, the probe can be easily cleaned by running a rod down the probe.

Heated Filter

The filter is housed in a weather-proof case which is mounted on the sampling port. It uses an oversize re-usable filter cartridge for removing fine particulate down to 1 um. The cartridge can be easily removed for maintenance. The element can be cleaned by reverse-blowing air or steam. The over-sized filter also serves as a drop out and evaporation chamber for any liquid mist present. It is heated to a temperature above typical acid boiling point (e.g. HCl, H2SO4 , etc.). With a large particulate collection volume and filtering surface, it allows for extended operation before maintenance is required.

Calibration Gas Inlet

A calibration gas inlet is provided upstream of the heated filter. This is an essential tool for complete system calibration and performance evaluation. Calibration gas standards can be introduced into the system near the sampling point. The calibration gas is exposed to the same flow path and conditioning as the sample gas. Any leak, contamination or problems in the sampling path can be identified.

Electrical Module

The module houses the temperature and blow-back control electronics. Connection strips are provided for easy electrical and blow-back status signal connections. The module is designed for outdoor applications and is weather-proof.

Specification

Gas Sample Conditions
Flowrate
0 - 15 L/min
Temperature
-58 to 2,100 oC), higher temperature available
Pressure
Positive or negative
Moisture Content
0 to 100% condensing or non-condensing
Corrosive Component
HCl, H2SO4, NH3 and other vapours (with optional scrubber)
Sampling Probe Material Stainless Steel. Teflon coated, Hastelloy, ceramic, and others optional
Filter
Housing
Stainless Steel, heated, Teflon wetted parts optional
Filter Element
Re-useable, 1 um
Blowback
Control
Build-in controller and external trigger (contact closure or PLC). Intelligent Controller (filter loading sensitive) optional
Duration & Frequency
10 to 50 psig. Higher pressures optional
Medium
Compressed air, steam optional
Compressed Air Supply
60 - 100 psig, 0.5 CFM
Compressed Air Quality
Dry, oiless, filtered
Connection Fitting
3/8" compression fitting
Status Signal Output
Digital (dry contact), 60 VDC or 125 VAC 0.3 A. For blow-back period identification
Mounting 3" ANSI 150 lb flange standard, 4" available, others optional
Max. Heating Sample Line Size 4" OD, other size available
Sample Gas Connection 1/4" compression fitting
Calibration Gas Connection 1/4" compression fitting, internal. External connection available
Electrical & Signal Connection Knockout 1/2" conduit
Housing Weather-proof
Power Supply 115 / 220 / 240 VAC 60 / 50 Hz 700 VA
Area Classification General purpose. Hazardous area classification version optional
Operating Environment -40 to 70 oC (-40 to 158 oF), vibration free, non-corrosive (higher operating temperature available as option)

Options

In-Stack Pre-FilterTeflon Wetted PartsIntelligent Controller
Filter Blockage Sensor with AlarmSampling PumpRadiant Heat Shield
Water Cooled Sampling ProbeSample CondenserSample Dilution Kit
Hazardous Area VersionSteam Blow-Back